HPLC SOP

What is the correct order of purging a pump?
Open drain valve -> Press Purge -> Close the Drain Valve
Press Purge -> Open drain valve -> Close Drain valve
Close Drain valve -> Press Purge -> Open Drain valve
Why is the maximum pressure limit important?
It changes the behaviour of the analytes.
Too high of a backpressure can be safety issue
What function does the mobile phase play in a HPLC?
It removes dissolved gases in the solvent which can interfere with operation of the HPLC
Supplies the force necessary to provide a constant flow rate regardless of the resistance met in the system.
Interacts with the sample to retain the compound in the column.
Moves the sample through the column.
When I start a new data run, a prompt shows up on the LabSolutions Software telling me to press "Start". Do I follow the instruction?
Yes
No
What do I have to look out for when equilibrating the column?
Constant pressure
10 column volumes have passed
Desired temperature has been reached
Check for leaks
When I am doing Peak Integration of my data run. The LabSolutions software can detect more than 50 peaks! I am only expecting 7 compounds. How do I fix this?
Rerun the sample
Calculate by Height and change the minimum height to 10000 AND change slope to 10000
Change slope to 10000 only.
Change min/area height to 10000.
I want to save a chromatogram and be able to modify it. How do I do that? Tick the options that apply
This is called a metafile
Right click on chromatogram and click "Copy as Metafile"
In a Powerpoint slide, click the arrow under Paste under the Home toolbar.
Click Paste Special then a window should appear named ‘Paste Special’. Then select Picture (Enhanced Metafile).
Ungroup the image twice.
Where do I change %B and the Oven temperature respectively?
LC Time Prog, Detector
Data Acquisition, Detector
Pump, Column Oven
Detector, Column Oven
What do I expect when I increase the oven temperature?
Increase in backpressure
Increased retention time and increase in backpressure
Decrease in backpressure and decreased retention times
Decrease in retention times.
Which of these is not a module of a HPLC system?
Detector
Oven
Column
Pump
Mobile phase
Detector
Injector System
Backpressure
Which of these responses are not calculated?
Resolution
Retention Time
Retention Factor
Selectivity
True or False: Manual HPLCs have a UV VIS detector and UPLCs have a PDA detector.
True
False