Posterior expandables & CREO

RISE: The adjustable trial should always be cleaned in the closed position.
True
False
ALTERA: The Altera Inserter Sleeve is assembled into the gold knurled portion (Inserter Base) by turning clockwise in the Inserter Base until finger thight.
True
False
ALTERA: What two implant assembly components have an etched black arrow on it that should point towards each other (aligned) upon insertion?
Inserter Base & Inserter Threaded Rod
Inserter Base & Inserter Sleeve
Inserter Threaded Rod & Inserter Sleeve
Inserter Base & Inserter Driver Shaft
ALTERA: To detach the implant from the Altera Inserter Assembly, the surgeon must…
Remove the Palm Handle and turn the golden knob counterclckwise
Remove the Palm Handle and turn the Threaded Rod clockwise
Remove the Palm Handle and rotate the Threaded Rod counterclockwise
Remove the Palm Handle and turn the golden knob clockwise
ALTERA: After thigthening the implant (finger thight) onto the Threaded Rod, what recommended instrument is used to ensure the implant is secure?
The Spanner Wrench
Removal Threaded Rod
Inserter Sleeve
Inserter Base
ALTERA: The starting height of the 15° lordotic implants is 9mm.
True
False
CALIBER: How can the maximum expansion of the implant, after implantation, be visually verified?
Through hole in the middle of the implant in lateral xray
The etched ring on the Torque Limited Instrument is flush with the Inserter Handle
The radiographic markers are alignedvertically in lateral xray
RISE: What is the visual confirmation to ensure the Inserter Fork and the Inserter Tube are correctly assembled?
The proximal end of shaft should not be visible through the visualization hole in the Inserter Base
The black arrows should be aligned
The proximal end of shaft should be visible, halfway through the visualization hole in the Inserter Base
The distal ends of Inserter Fork und Inserter Tube should be flush
ALTERA: What are the three length options?
22mm, 26mm, 31mm
26mm, 30mm, 26mm
26mm, 31mm, 36mm
26mm, 31mm, 39mm
ALTERA: What are the two benefits compared to a static banana cage?
Packability
Ease of placement
Sagittal Balance
Alignment
ALTERA: What are the radiographic landmarks to confirm full articulation in the lateral view?
€žHourglass“ shape posterior to midline
Notch along posterior edge of the implant
Thru hole in the middle of the implant
Expanded implant
Notch along anterior edge of the implant
€žHourglass“ shape anterior to midline
ALTERA: What are the radiographic landmarks to confirm full articulation the AP view?
Notch along anterior edge of the implant
Thru hole in the middle of the implant
Notch along posterior edge of the implant
Expanded implant
ALTERA: What Sets are needed for a case?
Altera Trials
Altera Instruments
Posterior Disc Prep I
Signature Instruments
Posterior Disc Prep II
Altera Implants
RISE: How many Nm is the torque driver?
1.5Nm
2.5Nm
2Nm
1.75Nm
RISE is intended for what levels?
L1-S1
L5-S1
L1-L5
L2-S1
RISE offers how many mm of expansion?
4mm
8mm
7mm
3mm
RISE: What are the lordotic options?
7°
15°
6°
10°
0°
4°
RISE: What are the width options?
10mm, 12mm, 20mm
8mm, 10mm, 12mm
7mm, 9mm, 11mm
7mm, 10mm, 12mm
RISE: You can use the Caliber torque driver.
True
False
CALIBER is intended for for use in a max of ___ levels.
1
2
3
4
CALIBER: Which widths are available?
10mm, 12mm, 18mm
8mm, 10mm
10mm, 12mm
8mm, 10mm, 12mm
CALIBER: What lengths are available?
22mm, 26mm, 30mm
20mm, 26mm, 30mm
18mm, 22mm, 26mm
26mm, 31mm, 36mm
CALIBER: How many degrees is the largest lordotic option?
15°
10°
20°
21°
CALIBER: it is okay to rotate the implant when attached to the Inserter.
True
False
CALIBER: on a 9-13mm implant, ___ turns on the torque limiting driver indicates 1mm of expansion.
2
2.5
1.75
1.5
CALIBER: The purpose of the torque limiting feature of the trial and implant is to protect the integrity of the instrument/implant not the endplate.
True
False
CALIBER: It is okay to impact on the torque limiting driver when placed in the insertion handle.
True
False
ALTERA: How many independent Clinical studies are available on ALTERA?
3
2
1
0
ALTERA: What sagittal profiles are offered?
15°
18°
10°
8°
0°
4°
ALTERA: How many mm of expansion?
7mm
4mm
8mm
6mm
ALTERA: What are the starting heights?
7mm
9mm
10mm
8mm
12mm
11mm
CREO AMP Threaded/ Screw Head Insertion: For correct head attachment, the UNLOCKED position shows a high saddle and visual silver ring below.
True
False
CREO AMP Threaded: Out of the six reduction options for the Creo Amp Threaded System, which instrument reduces 30mm?
Tower Reducer
Geared Reducer, Overhead
Geared Reducer, Pistol Grip
Reduction Clip
CREO: What are minimum sets to order for a Creo Degen case?
Creo Deformity Implants & Instruments
Creo Core Instrument
Creo large diameter
Creo Degen Implants
Creo Expanded Implant Set
Creo Degen Instruments
CREO: The taps are undersized by how many ___mm?
0.1mm
1.1mm
0.2mm
0.3mm
CREO: Tapping is required.
True
False
CREO AMP threaded uses the black torque limiting handle with 8Nm.
True
False
What´s the name of the handle for the Inserter Assembly for ALTERA?
MIS Handle
Inserter Handle
Torque Limiting Palm Handle
QC Handle
What is the starting height for a 15° RISE cage?
7mm
8mm
9mm
10mm
ALTERA: You always need to use the Spanner Wrench to securely attach the Implant to the Implant Assembly Holder.
True
False
ALTERA: The Implant Assembly Holder corresponds to the height of the implant.
True
False
RISE & CALIBER expand up to 7mm.
True
False
CREO AMP threaded accepts which rod diameter(s)?
4.75mm
5.5mm
6.35mm
6.0mm
CREO AMP 5.5 accepts which rod diameter(s)?
4.75mm
5.5mm
6.35mm
6.0mm
CREO AMP 5.5 final tightens with the blue handle 5.5NM.
True
False
I can use the CREO AMP 5.5 Reduction Tower for a CREO AMP threaded screw.
True
False
ALTERA: What are the preferred/recommended angles for insertion?
10°, 15°, 10°
15°, 10°, 15°
10°, 15°, 0°
15°, 0°, 15°
What are differences between CREO AMP Threaded and CREO AMP 5.5?
Screw Diameters
Rod Acceptance
Torque limit/ final tightening
Color of tulips
Locking Caps
Reduction options
Screw Driver features (e.g. Hexalobe/Trialobe,…)
Cross Connectors
{"name":"Posterior expandables & CREO", "url":"https://www.quiz-maker.com/QPREVIEW","txt":"RISE: The adjustable trial should always be cleaned in the closed position., ALTERA: The Altera Inserter Sleeve is assembled into the gold knurled portion (Inserter Base) by turning clockwise in the Inserter Base until finger thight., ALTERA: What two implant assembly components have an etched black arrow on it that should point towards each other (aligned) upon insertion?","img":"https://www.quiz-maker.com/3012/images/ogquiz.png"}