Chapter 14: High Temp Sterilization

When combining loads, hard goods should be placed on the top shelves to allow for more efficient removal of the condensate.
True
False
Three of the main phases of a terminal steam sterilizer cycle are _____________, ____________ and _______________.
conditioning, exposure and exhaust
Pre-heat, Sterilization, Vent
Rinse, Sanitize, Dry
Load, Run, Unload
The steam sterilization process can be affected by the design of the medical device being sterilized.
True
False
Central Service technicians need to understand the anatomy of a steam sterilizer to ____________________.
understand how the sterilizer operates
Increase patient satisfaction
Maximize hospital revenue
Ensure proper sterilization of instruments
The weakest part of a steam sterilizer is the __________.
Door
Pressure gauge
Sterilizer chamber
Drainage Mechanism
The most common reason for steam sterilization failure is ______________.
lack of steam contact with the instrument
Clogged drain strainer
inadequate temperature
All of the above
The coolest place in a steam sterilizer is the _____________.
thermostatic trap/Drain
Chamber
Water reservoir
Exhaust valve
Steam flush pressure pulse sterilizers are a type of gravity sterilizers.
True
False
Factors that can cause sterilant contact failure with the instrument are ________________ and a _______________.
crowded loads / clogged drain strainer
Over sterilization time; discolored instrument surface
Under sterilization time; chipped instrument surface
Proper sterilant concentration; rusty instrument surface
One of the most frequent causes of a clogged drain screen is ___________.
Excessive use of soap
tape
Inadequate cleaning of instruments
The use of non-sterile instruments
After sterilization, the load contents may take two hours or more to cool.
True
False
Items with a standard steam sterilization cycle recommended by the manufacturer can be damaged if run in an extended cycle.
True
False
Immediate use steam sterilization is the process to sterilize trays for future use.
True
False
Wood products should be sterilized in an extended steam cycle.
True
False
When water is seen on the outside of a pack after sterilization the pack is considered safe to use if all other packs in the load are dry.
False
True
Packs that are improperly packaged or loaded on the sterilizer cart frequently become wet packs.
True
False
Peel pouches should be placed ________________ for sterilization.
on edge, paper to plastic
Flat on their side
Flat, Folded in half
Tightly packed with other items
0
{"name":"Chapter 14: High Temp Sterilization", "url":"https://www.quiz-maker.com/QJ3T5CCCU","txt":"When combining loads, hard goods should be placed on the top shelves to allow for more efficient removal of the condensate., The steam sterilization process can be affected by the design of the medical device being sterilized., Central Service technicians need to understand the anatomy of a steam sterilizer to ____________________.","img":"https://www.quiz-maker.com/3012/CDN/87-4242045/thesterileprocess-3-3-in-.png?sz=1200-00000021501000005300"}