CCHM311 Prelim Practice
Capable of providing a single test result on approximately 40 samples per hour.
Discrete Analyzer
Centrifugal Analyzer
Continuous Flow Analyzer
Automation
Following are not the disadvantages of automation, EXCEPT:
Minimized Variability
Initial Costs
Reduced Consumable costs
Reduced Sample Volumes
All samples are loaded at the same time.
Batch testing
Parallel testing
Random access testing
Sequential testing
One specimen with more than one test is analyzed
Batch testing
Parallel testing
Random access testing
Sequential testing
Able to perform individual test or panels.
Batch testing
Parallel testing
Random access testing
Sequential testing
System other than manufacturer's reagents can be utilized for measurement
Pneumatic Tube System
Closed Reagent System
Dry chemistry slide
Open Reagent System
Transports specimen quickly from one location to another
Closed Reagent System
Dry chemistry slide
Throughput
Pneumatic Tube System
A multilayered film technology use by automated analyzers
Closed Reagent System
Pneumatic Tube System
Automation
Dry chemistry slide
A mechanized device that automatically dips into a sample cup and aspirates a portion of liquid.
Probe
Closed Reagent System
Rotor
Automation
Maximum number of test generated per hour
Throughput
Turnaround
Standalone
Modular work cell
At least two analytical modules supported by one sample and reagent processing and delivery system
Multiple Platform
Modular work cell
Integrated Modular System
Closed tube sampling
It is also known as a "run-based" system which cannot be interrupted once the batch-testing measurement has started.
Continuous Flow
Discrete analysis
Centrifugal analysis
Automation
Have the capability of running multiple tests one sample at a time.
Continuous Flow
Discrete analysis
Centrifugal analysis
Automation
Time considered for a ___ is 20 seconds to 1 minute.
Kinetic Reaction
End-Point Reaction
Measure the total amount of analytes that participate in the reaction
Kinetic Reaction
End-Point Reaction
Ehrlich Reaction
Reflectance Photometry
Commonly utilized sample in the Clinical Chemistry Section
Heparinized Plasma
EDTA Plasma
Serum
Citrated Plasma
Lt is a special alumina-silicate glass that has been strengthened chemically than thermally; six times stronger than borosilicate.
Soft Glass
Flint Glass
Corex
Vycor
It is utilized for high thermal, drastic heat shock, and extreme chemical treatment with acids (except hydrofluoric) and dilute alkali; it can be heated to 900°C.
Soft Glass
Flint Glass
Corex
Vycor
Its thermal resistance is less as compared to borosilicate glass.
Soft Glass
Flint Glass
Corex
Vycor
This type of pipette usually has the greatest degree of accuracy and precision and should be used when diluting standards, calibrators, or quality control material
Pasteur Pipette
Volumetric Pipette
Ostwald Folin
Long-Levy pipettes
It has continuous etched rings on top or near the mouthpiece of the pipette and exact volume is obtained when the last drop is blown out.
Blow-out Pipette
Self-draining Pipette
It operates by moving the piston in the pipette tip or barrel, much like a hypodermic syringe.
Air Displacement Pipette
Dispenser/Dilutor Pipette
Positive Displacement Pipette
Volumetric Pipette
It relies on piston for suction creation to draw the sample into a disposable tip
Air Displacement Pipette
Dispenser/Dilutor Pipette
Positive Displacement Pipette
Volumetric Pipette
calibrating medium for TD pipettes
Mercury
Distilled water
Deionized Water
Strobe light
Centrifuge calibrated every?
1 month
2 weeks
4 months
3 months
Produces a tightly packed, flat sediment surface.
Horizontal-head Centrifuge
Swinging-bucket Centrifuge
Angle-head Centrifuge
Ultracentrifuge
1 of the choice is correct
All are correct
2 Choices are correct
Suitable for extreme temperature treatment (-70 °C to 200° C)
Vycor
Teflon
Corex
Flint Glass
It is approved for human consumption (not injurious to individuals) but may not be applicable for laboratory analysis.
USP
Ultra Pure Reagent
Analytical Grade Reagent
Commercial Grade
A solution in which there is an excess of undissolved solute particles
Dilute solution
Concentrated solution
Saturated solution
Super saturated solution
Unsaturated solution
Calibrator for Centrifuge
Distilled Water
Type II reagent water
Mercury
Strobe light
Study of phenomenon of light adsorption by molecules in solution
Photometry
Spectrophotometry
Fluoriscence
AAS
Isolates light within a narrow region of the spectrum.
Exit Slit
Entrance Slit
Monochromator
Cuvette
This utilized a flameless dehydrates and atomizes a sample using a graphite capillary surface
AAS
FEP
Spectroscopy
Mass Spectroscopy
Distance traveled by one complete wave cycle
Bandpass
Trough
Wavelength
Peak
CUVETTE for Visible incandescent
Tungsten
Tungsten Iodide
Quartz
Borosilicate
Color transmitted by Blue
Orange
Red
Red-Orange
Green
575-600nm
Blue
Orange
Yellow
Green
Emits radiation that changes in intensity
Line Source
Continuum Source
Emits limited radiation and Wavelength
Line Source
Continuum Source
Bandpass
Allows only a narrow fraction of the spectrum to reach the sample cuvette
Exit Slit
Monochromator
Entrance Slit
PMT
Used for measurement of solution requiring visible and ultraviolet spectra
Alumina Silica Glass
Quartz
Borosilicate
Soft Glass
Basic photodetector for detecting and measuring radiation in the visible region
Barrier layer cell
PMT
Photodiode
Diffraction grating
Not sensitive as PMT but with excellent linearity
Barrier layer cell
PMT
Photodiode
Diffraction grating
Chemical reaction produces colored substance that absorbs light of a specific wavelength
Spectrophotometer
Photometry
Fluorescense
Chromatography
With 2 photodetectors, for the sample beam and reference beam
Double-beam in space
Double-beam in time
Element is not excited by merely dissociated from its chemical bond and place in an unionized, unexcited ground state.
FEP
Spectroscopy
AAS
MS
It is based on the fragmentation and ionization of molecules using a suitable source of energy
FEP
Spectroscopy
AAS
MS
Measures reduction in light transmission by particles in suspension and light is measured at angle from light source
Turbidimetry
Nephelometry
Fluorometry
Chemiluminiscence
Migration of charge particles in an electric field.
Fluorometry
Chromatography
Electrophoresis
FEP
Gold standard in drug testing
GC-MS
TLC
HPLC
Liquid Chromatography
Has a net charge that can be either positive or negative depending on PH conditions
Endosmosis
Iontophoresis
Zone Electrophoresis
Amphoteric
Is the movement of buffer ions and solvent relative to the fixed support
Endosmosis
Iontophoresis
Zone Electrophoresis
Amphoteric
Migration of small ions
Endosmosis
Iontophoresis
Zone Electrophoresis
Amphoteric
Separates by electrical charge
Cellulose acetate
Agarose Gel
Polyacrylamide Gel
Separates proteins into 20 fractions used to study iso enzymes
Cellulose acetate
Agarose Gel
Polyacrylamide Gel
It scans and quantitates electrophoretic pattern.
Densitometry
Coomasie Blue
Photometry
Gold/Silver Stain
Very sensitive even to nanogram quantities of proteins
Densitometry
Coomasie Blue
Photometry
Gold/Silver Stain
Ideal for separating proteins of identical sizes but with different net charges
Direct ISE
Indirect ISE
Ion selective Electrode
Isoelectric focusing
Ideal temperatures higher than 60 degrees Celsius and react with more sample
Silver-Silver Chloride
Saturated Calomel Electrode
PCO2 level in the sample is determined indirectly by sensing the pH change in this solution
Glass pH electrode
Severinghaus electrode
Not affected by hyperproteinemia and hyperlipidemia
DIRECT ISE
INDIRECT ISE
Measurement of the current flow produced by an oxidation reaction
Potentiometry
Chromatography
Electrophoresis
Amperometry
Amount of increase in current proportional to the concentration of analyte
Potentiometry
Electrophoresis
Polarography
Amperometry
It is the measurement of current after which a potential is applied to an electrochemical cell
Potentiometry
Electrophoresis
Polarography
Voltammetry
Separation of solute is based on relative solubility in an organic solvent and an aqueous solvent
Adsorption Chromatography
Steric Exclusion
Partition Chromatography
Ion-Exchange Chromatography
Is based on the competition between the sample and the mobile phase for sites on the solid stationary phase.
Adsorption Chromatography
Steric Exclusion
Partition Chromatography
Ion-Exchange Chromatography
A semiquantitative drug screening test.
GC-MS
TLC
MS
FEP
Uses pressure for fast separations
TLC
GC-MS
MS
HPLC
What is the common source of light for the UV region
Deuterium Lamp
Tungsten Iodide lamp
Xenon lamp
Mercury Arc
In the SI unit What is the base quantity of time?
Minutes
Hours
Seconds
Days
To convert INCHES into CENTIMETERS multiply by?
3.14
2.34
0.39
2.54
To convert POUNDS into KILOGRAMS multiply by?
0.49
0.14
0.39
0..45
Instrument able to perform tests from at least two disciplines
Multiple Platform
Stand alone
Modular work cell
Closed sample tubing
Atleast two instrument from a single discipline with one controller
Multiple Platform
Stand alone
Modular work cell
Closed sample tubing
Kaya pa ba?
Kaya pa
Kakayanin
Siguro
Di na
Suko na
{"name":"CCHM311 Prelim Practice", "url":"https://www.quiz-maker.com/QPREVIEW","txt":"Capable of providing a single test result on approximately 40 samples per hour., Following are not the disadvantages of automation, EXCEPT:, All samples are loaded at the same time.","img":"https://www.quiz-maker.com/3012/images/ogquiz.png"}
More Quizzes
Square Roots Kitchen (7/18)
520
3eso geo
1059
OCI and Rodeo Trivia
1260
Test 1-10
52260
Ennagrammi testi
4224
Getting to Know You...Again!
241228
How Well Do You Know Addison?
1165
The history of Guapa Golf Tours
201020
How Well do You Know Kirana and Anin?
1169
Skin Cancer winx club!
10511
Shadow the Hedgehog Ending OR Megadeth Song
40200
How professional are you?
740